Conmed Aspen Excalibur Plus is an isolated generator that allows surgeons to adjust the power settings from within the surgical area using a standard ES pencil. It has cut, bipolar, coag, and blend modes. It features a Controlled Output Voltage which maintains the minimum voltage necessary for smooth, scalpel-like cutting. Aspen excalibur plus service manual, adobe after effects cc guida alluso, ratfked the true story behind the secret plan to steal americas democracy, essentials of musculoskeletal care 4th edition, liverpool academy coaches manual enrych, contax 159mm camera repair manual ebooks, cat.
- Videos
- News
DESCRIPTION
The ConMed Excalibur Plus PC is the most powerful isolated ESU available in its class.The ConMed Excalibur Plus PC is the most powerful isolated ESU available in its class. It allows the surgeon to adjust the power settings from the surgical field with a standard ES pencil. The Excalibur Plus PC patented Controlled Output Voltage in CUT maintains minimum necessary voltage for smooth cutting through various impedance tissues with minimal thermal damage.
FORUMSView All (3)
Ask a New Question1Reply6 days ago | 6 days agoElectrosurgical unit I want service manual pleaseReply |
-flmd 3 years ago | 3 years agoCOnmed- Excalibur Plus Is there a new version of this? i got the 1998 version. Please advise. Thank youReply |
1Reply7 years ago | 7 years agorequire service manual for excalibur plus Dear sir I require the service manual with schematics for Conmed excalibur plus. Can you help me with one Reply |
DOCUMENTS / MANUALSView All
SERVICE COMPANIESView All Electrosurgical Unit (ESU) Companies
FEATURES
- 10,000V PTP in Spray COAG.
- It is the only ESU that allows the surgeon to adjust power settings from the surgical field with a standard ES pencil.
- Excalibur Plus PC Patented Controlled Output Voltage in CUT maintains minimum necessary voltage for smooth cutting through various impedance tissues with minimal thermal damage.
- The Excalibur Plus PC offers a scalpel like cut and three blend modes.
Additional Specifications
Pure Cut Maximum Peak to Peak Voltage is 2000 Volts.
Blend 1 - Maximum Peak to Peak Voltage is 2000 Volts.
Blend 2 - Peak voltage of 2200.
Blend 3 - The Excalibur Plus PC comes equiped with the maximum Peak to Peak Voltage
of 2400.
Aspen Excalibur Plus Service Manual 2016
- Documents
- Videos
- News
DESCRIPTION
High Power Digital ESUAspen Excalibur Plus Service Manual 2017
The Conmed Aspen Excalibur is a high power digital electrosurgical unit and features unsurpassed fulguration enhanced bipolar performance with less forcep sticking and simultaneous activiation of 2 pencils.
FORUMSView All (1)
Ask a New Question1Reply7 years ago | 7 years agocheckout procedures anyone have a manual for this unit? Reply |
SERVICE COMPANIESView All Electrosurgical Unit (ESU) Companies
FEATURES
- Power Plus and PC Plus Models available
- Bipolar and Monopolar
- Multi-Purpose Electrosurgical Generator
- Microprocessor Control provides the user with Safety and Control
- Bipolar and Monopolar cutting and Coagulation
- 300 Watts Output - Cut
- Programmable Digital Display
- Reusable Groundplate
- Biopolar Footswitch (Optional)
- Rolling Carts (Option)
Aspen Excalibur Plus Service Manual Online
SPECIFICATIONS
OutPut Frequency | 416.7 kHz Cut. ±0.01%. Coag: 540 kHz ±10%. Bipolar: 1.05 MHz |
Height | 6.25 in |
Length | 22.75 in |
Weight | 40 lbs |
Width | 15.9 in |
Additional Specifications
Aspen Excalibur Plus Service Manual Download
Pulse Repetition Frequency: Blend, spray and Bipolar Coag: 20 kHz ± 0.1%. Standard Coag: 20 kHz ±0.1%
Input Voltage: 104-127 VRMS. 60 Hz calibrated at 120V
Input Current: Idle: 0.8A. Pure Cut at 300: 6.5 A. Coag at 120: 2.8 A
Line Frequency Leakage: Patient connected to neutral < 10 ΜA. Chassis to neutral < 25 ΜA
RF Resolution: 99% Minimum
Power Display Accuracy: Within 10%, or 3 watts at rated load